    Table 1. Formaldehyde Method Characteristics


    *60 seconds in high accuracy mode and 8 seconds in low accuracy mode. For this study, the
    PPM was in high accuracy mode.
    Description of the Direct-Reading Instruments and Laboratory-Based Method
    The RKI Instruments Model FP-30 uses photoelectric photometry with colorimetric detection
    tabs to measure formaldehyde in air (RKI Instruments Inc., Hayward, California) (Figure A-1 in
    Appendix). The RKI Instruments Model FP-30 draws in air with an internal pump and uses a
    microprocessor to control sample flow rate. The instrument uses detection tabs to sample
    airborne formaldehyde. Depending on the desired detection range, a particular tab number is
    used during sampling. For this study, number 009 detection tabs were used to measure
    formaldehyde concentrations ranging from 0 to 1.00 ppm in a 15-minute sampling time. The
    instruction manual states the RKI Instruments Model FP-30 is capable of measuring
    formaldehyde concentrations ranging from 0 to 1.00 ppm with a resolution of 0.01 ppm.
    According to the manufacturer, measurements are accurate to ± 10% when air temperature is
    between 14ºF and 104ºF and the relative humidity is below 90%.